3.4 Setting Configuration 2 – Input Channel Settings
60
[
°
C]
1000
400
0
-200
-1000
When the Thermocouple Measurement Range is Exceeded or
the Allowable Measurement Range of Each Range is Signifi-
cantly Exceeded
• Example: Using a type T thermocouple with range -200°C to
+400°C
When measurements are outside of the range of the thermocou-
ple, the waveform saturates and clings to the top of the screen as
shown at the right.
If the allowable measurement range of each range is significantly
exceeded, “OVER” or “-OVER” is displayed for the numerical val-
ue or cursor value and the calculated value or saved data is han-
dled as a full scale value.
Refer to
“LR8432-20 Measurement Values” (p. A13).
About Burn Out Detection
Broken (disconnected) thermocouple wires are detected using a tiny (300 nA) sensing
current. The thermal equivalent of the thermoelectromotive force (thermocouple resis-
tance (
Ω
) times the 300 nA sensing current) is added to the measured value. However,
measurement errors can occur if the thermocouple wire is long or if the wire itself has a
large resistance, in which cases broken wire detection should be disabled (set to Off).
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......