![Hioki LR8432-20 Instruction Manual Download Page 72](http://html1.mh-extra.com/html/hioki/lr8432-20/lr8432-20_instruction-manual_2137139072.webp)
3.4 Setting Configuration 2 – Input Channel Settings
64
Configure these settings for each pulse channel on which you will be counting
pulses corresponding to revolutions, such as output from a rotary encoder or
tachometer. The revolution measurement is obtained by counting the number
of pulses input per second. The scaling function can be used to display the
results of converting measured values to physical properties of the measure-
ment target (p. 69).
Revolution Measurement Settings
Change by 1
Change by 10
Detect-
ing level
L
H
1
Select a channel (
P 1
to
P 4
) for the
[Chan-
nel]
setting contents.
2
To enable measurement on the channel,
select On to display the check mark (
).
3
Select
[Revolve]
for the
[Input]
(input type)
setting.
4
Set the
[Pulse/Rev]
setting to the number of
pulses per revolution.
Setting options:
1
∗
to
1000
(
∗
default setting)
The range is automatically determined by this set-
ting.
5
Select the basis of measurement from the
[Slope]
setting.
Setting options:(
∗
default setting)
↑
∗
Counts the number of revolutions based on
L - H transitions.
↓
Counts the number of revolutions based on
H - L transitions.
5000 r/s (n=1)
1 2
: On
: Off
3
4
5
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......