![Hioki LR8432-20 Instruction Manual Download Page 151](http://html1.mh-extra.com/html/hioki/lr8432-20/lr8432-20_instruction-manual_2137139151.webp)
7.3 Performing Waveform Calculation
143
3
7
C
h
ap
te
r 7
N
u
m
e
ri
cal
/W
avef
orm
C
a
lc
u
lat
ions
The calculation formula can be entered in the
[Conv/Calc]
screen.
The W1 calculation formula can be copied to other channels. (p. 74)
1
Select the
[Conv/Calc]
screen.
2
Select
[Calc]
.
3
Select the waveform calculation formula
for the channel (
W1 to W10
).
The dialog box is displayed.
4
Select the calculation type and set nec-
essary items.
Increases/decreases by 1, c/-.
Moves the digit.
Coefficient settings dialog box
2
3
Description
Calculation type
Setting item
Moving average
Enter/select CHA, number of points, and unit and press the
ENTER
key. (The upper/lower limit values for the selected waveform calculation
are the same as for the target calculation channel.)
Average
Enter/select CHA, start reset, reset time, and unit and press the
EN-
TER
key. (The upper/lower limit values for the selected waveform cal-
culation are the same as for the target calculation channel.)
Sum
Enter/select CHA, start reset, reset time, and unit and press the
EN-
TER
key.
Heat flow rate
Enter/select CH, temperature CH1, temperature CH2, and unit and
press the
ENTER
key.
This is a reference formula to calculate the heat flow rate.
Four operations
Enter/select coefficient a, CHA, calculation, coefficient b, CHB, coeffi-
cient c, and unit and press the
ENTER
key.
Reset timing
Select whether or not to reset the calculation at the point of a trigger
event. (Selection
OFF, trigger position)
Reset time
Select whether or not to reset the calculation at the specified time.
(Selection
OFF, ON, scheduled time)
When ON is set, the calculation is reset at the specified reset time in-
terval. When Ref Time is set, the calculation is reset at the reset time
interval beginning at the specified reference time.
4
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...