3.5 Display Settings
71
3
3
C
h
apt
er 3
Set
ting
be
fo
re M
easurement
Setting Pulse Channel (Integration) Scaling
Integrated pulse values can be converted for display as physical units of the mea-
surement parameter (such as Wh or VA).
Pulse output devices determine the amplitude per pulse in physical units, or the num-
ber of pulses per fundamental unit (for example, per kWh, liter or m
3
).
1
For the
[Cond]
(condition) setting, select the
desired numerical notation.
Setting options:(
∗
default setting)
2
Enter the
[Unit]
name on the comment entry
screen.
Refer to
3
Specify the physical units per pulse or the
number of pulses per fundamental unit (for
example, if 1 c = 1 pulse, select
[1 pulse]
).
The settings for physical units per pulse and
pulses per fundamental unit are interlinked.
Dec
Displays converted values using fixed-point
notation
Exp
Displays converted values using floating-
point notation
Off*
Scaling is disabled.
1
3
2
Using fixed-point
notation
Using floating-point
notation
Scaling Setting Example
For integration with a 50,000 pulse/kWh watt-hour meter:
Cond: Dec
Units: k Wh
1 kwh=50000 pulse
For integration with a 10 liter/pulse flow meter:
Cond: Dec
Units: L
1 pulse=10
L
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......