Inspection Procedures
3-9
P/N 108755-US, Manual revision: AA
Ribbon cable requires replacement.
Replace ribbon cable. See Chapter 13,
Cap Assembly Removal on page 13-3.
Audible alert fails to sound
with power switchon and
shroud depressed.
Audible alert failure.
See Chapter 12, Audible Alert Failure
on page 12-2.
Damaged or disconnected wires to
audible alert.
Replace or repair wires.
Other component failure.
Follow malfunction isolation section
for suspected component.
Display or indicator lights do
not illuminate.
LED burnt out.
Replace head board. See Chapter 13,
Head Board Replacement on page 13-
5
Keypad button does not acti-
vate properly.
Keypad malfunction.
Replace keypad. See Chapter 13, Key-
pad Replacement on page 13-6.
Display reads "gibberish" or
unintelligible words.
Head board or body board malfunc-
tion.
Replace head board. See Chapter 13,
Head Board Replacement on page 13-
5 and Chapter 9, Body Board Replace-
ment on page 9-2.
Table 3-1, General Troubleshooting Guide
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......