5-2
Strain Gauge
P/N 108755-US, Manual revision: AA
S
TRAIN
G
AUGE
I
NSPECTION
AND
R
EPLACEMENT
Caution:
Always wear an ESD strap when working on the Shaker.
Figure 5-1, Body Board and Power Supply Assembly Removal
Figure 5-2, Nut Plate and Chassis Plate Mounting Screws
1.
With the base plate and shroud removed from the Shaker, remove the two
screws securing the chassis plate to the bracing bar.
Motor
Chassis Plate
Mounting Screw
Power Supply Assembly
(Partial View)
Chassis Plate
Mounting
Screws
Two screws that hold body board
and power supply assemblies and
assembly chassis plate to the
nut plate
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......