Link Assembly
6-3
P/N 108755-US, Manual revision: AA
L
INK
A
SSEMBLY
R
EMOVAL
Warning: Tension on the link assembly may cause it to spring forward when
released. Never place head or other body parts in front of link assembly during
removal or injury could occur.
Figure 6-2, Link Assembly Removal and Installation
1.
Remove the link assembly by detaching the shoulder bolt that secures the
Link M to the Crank M.
Note: The threads on the shoulder bolt are left-handed.
2.
Remove the two nuts and lock washers that secure the link assembly to
the sides of the base. Remove the washers from the left side and retain for
use.
Caution:
Nylon washers must be replaced on the
left
side only.
3.
Gently pull the link assembly away from the shaft and remove.
Link D
Link M
Shoulder Bolt
Crank M
Nut and
Lockwasher (2x)
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......