6-6
Link Assembly
P/N 108755-US, Manual revision: AA
4.
Using the Spring Tension Wrench (P/N 10400026) on the squared part of
the right side of the spring guide, turn the spring guide accordingly to even
the noise level.
5.
Tighten both nuts and turn on the Shaker.
6.
Repeat steps 2–4 until an optimum noise level has been achieved.
7.
Press the <*> to cycle shroud assembly, count the cycles for one minute
(CPM). The CPM should be 38cpm ±5 cycles.
Note: If the power supply PCBA is P/N 10403200, RV1 can be adjusted to set the
cpm. If p/n 10403600 is used, resistor R6 controls the cpm. Verify the body board
R3, R7 calibration, reference Chapter 5, Body Board Strain Gauge Amp Adjust-
ment Procedure on page 5-7.
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......