13-6
Cap Assembly
P/N 108755-US, Manual revision: AA
K
EYPAD
R
EPLACEMENT
1.
Pull head board out of cap assembly by removing the four mounting
screws.
2.
Unplug the keypad from the head board.
3.
Peel up the old keypad, ensuring that the residual adhesive is removed
from the face of the cap.
4.
Peel the backing off of a new keypad and center it on the cap. Thread the
keypad connector through the slot and plug it into the head board.
Caution:
Do not attempt to adjust keypad once it has been placed or the tactile
dome will be damaged and a new keypad will be required.