GSI Group - Laser Division
JK300HP/JK300P/JK125P
Pre-installation and User Manual
Issue 2.0
Page 10 of 172
Part No: 1EA300E01
1.8 Technical
Specification
This table provides general technical data on the JK300HP/JK300P/JK125P Series Laser
including the output specification and the services necessary for Laser operation.
Environment
JK300HP/JK300P/JK125P
Storage Environment
Temperature not to fall below 2°C (35°F) unless the
cooler has been drained.
Operating Environment
Max temperature:
40°C (104°F)
Min temperature:
5°C (40°F)
Humidity Specification
Max humidity:
95% RH at 20°C (68°C)
Max humidity:
50% RH at 40°C (104°F)
These figures indicate the maximum humidity at the
upper and typical temperature limits. Refer to Figure 1
for humidity figures within the temperature range.
Electrical Supply Requirements
JK300P
JK125P
JK300HP
Supply
Three-phase, three-wire plus earth (ground).
Line Voltages
380 - 415 V
±
10% at 50/60Hz
Voltage Tolerances
±6%
Supply Rating
11 kVA
8 kVA
15 kVA
Maximum Power Consumption
9.5 kW
6 kVA
13.7 kVA
Note:
This equipment is built to operate at 50Hz or 60Hz within the relevant supply voltage
range. This frequency is not interchangeable because a different Cooler Pump motor is used in
each case. Consult GSI Group - Laser Division if it is necessary to operate this equipment on a
different supply frequency.