Page 88 of 172
Part No: 1EA300E01
6.
Configurable threshold set too low.
7. Loose
connections.
8.
Faulty temperature sensor.
9.
Faulty Cooler Module.
7.5.10 Alarm 3n10: High Conductivity fault
Meaning:
The coolant conductivity has raised above a configurable threshold.
Possible causes:
1. Water
contamination.
2.
Cooler filters require changing.
3.
Configurable threshold set too low.
4. Loose
connections.
5.
Faulty conductivity sensor.
6.
Faulty Cooler Module.
7.5.11 Alarm 3n11: Low Coolant Flow
Meaning:
The coolant flow level has fallen below a configurable threshold.
Possible causes:
1.
Loss of coolant in the system.
2. Water
leak.
3.
Flow threshold set too high.
4.
Incorrect flow meter factor setting.
5. Loose
connections.
6. Faulty
sensor.
7.
Faulty Cooler Module.
7.5.12 Alarm 3n12: High Coolant Flow
Meaning:
The coolant flow level has raised above a configurable threshold.
Possible causes:
1.
Air in the cooling system.
2. Water
leak.
3.
Flow threshold set too low.
4.
Incorrect flow meter factor setting.
5. Loose
connections.
6. Faulty
sensor.
7.
Faulty Cooler Module.
7.5.13 Alarm 3n13: Temperature Sensor Fault
Meaning:
The coolant temperature sensor is faulty.