GSI Group - Laser Division
JK300HP/JK300P/JK125P
Issue 2.0
Pre-installation and User Manual
Part No: 1EA300E01
Page
97 of 172
7.10.12 Warning 1n12 : Shutter Multi shot or Single shot not allowed while
laser is off
Meaning:
A shutter multi pulse (Pulsed lasers) or single shot (CW lasers) was attempted when the
laser was not on. Turn the laser on, and retry the operation.
Reset on read.
7.11 Cooler Warnings (Zone 3)
7.11.1 Warning 3n01 : Low Temperature
Meaning:
The coolant temperature is below a configurable threshold.
Reset when coolant temperature raises above the configurable threshold.
7.11.2 Warning 3n02 : High Temperature
Meaning:
The coolant temperature is above a configurable threshold.
Reset when coolant temperature falls below the configurable threshold.
7.11.3 Warning 3n03 : Low Flow
Meaning:
The coolant flow is below a configurable threshold.
Reset when coolant flow raises above the configurable threshold.
7.11.4 Warning 3n04 : High Flow
Meaning:
The coolant flow is above a configurable threshold.
Reset when coolant flow falls below the configurable threshold.
7.11.5 Warning 3n05 : High Conductivity
Meaning:
The coolant conductivity is above a configurable threshold.
Reset when coolant conductivity falls below the configurable threshold (usually after a filter
change).
7.12 Power Supply Warnings (Zone 4)
7.12.1 Warning 4n01 : Disabled
Meaning:
The power supply has been disabled.
Reset when power supply enabled.