7.10.5 Warning 1n05 : Shutter Communications Timeout
Meaning:
A shutter open command was not received within 5 seconds of the previous command.
The shutter has closed due to the timeout.
Reset on read.
7.10.6 Warning 1n06 : Sector Over Range
Meaning:
The sector number received was over range. It was either over 20, or over the number of
active sectors configured in the pulse shape. Use the set entire pulse shape to add further
sectors to a pulse shape up to a maximum of 20 sectors.
Reset on read.
7.10.7 Warning 1n07 : Operating Envelope Over Range
Meaning:
The combination of pulse shape data exceeded the power supply working envelope.
Reset on read.
7.10.8 Warning 1n08 : High Output Trip
Meaning:
The laser output exceeded the configured output threshold limit.
Laser performed the specified action. Either continued, or entered standby.
Reset when laser restarts or enters on from standby.
7.10.9 Warning 1n09 : Low Output Trip
Meaning:
The laser output fell below the configured output threshold limit.
Laser performed the specified action. Either continued, or entered standby.
Reset when laser restarts or enters on from standby.
7.10.10
Warning 1n10 : Shot Limit Exceeded
Meaning:
The pulse laser shot threshold limit has been reached.
The lamps require changing to ensure specified laser output.
Reset when shot count is reset.
7.10.11
Warning 1n11 : Shutter Multi Pulse External Trigger Timeout
Meaning:
The shutter was open, in multi pulse mode, with external trigger enabled, but an external
trigger signal was not received for 30 seconds.
Reset on read.