Refer to Figure 23 for item identification.
Step Action
WARNING
ELECTRIC SHOCK HAZARDS EXIST.
ELECTRIC SHOCK CAN CAUSE SERIOUS OR FATAL INJURY.
Removal Procedure
1
Ensure Power is switched OFF.
2
Pull back the protective Cover (item 1) to expose the Focus Head (item 7).
3
Unscrew, but do not remove, the Clamping screw (item 6) to loosen the Retaining
Clamp.
4
Carefully and in a straight line, withdraw the Fibre Output Termination (item 2) from
the Focus Head (item 7) and remove the Focus Head.
5
Place the Focus Head (item 7) in protective packaging to prevent ingress of dirt or
moisture.
6
Install a protective Cover (not illustrated) over the Fibre Output Termination (item 2).
Installation Procedure
1
Remove protective Covers from the Focus Head (item 7) and the Fibre Output
Termination (item 2).
2
Inspect the Focus Head (item 7) for damage. Replace if necessary.
3
Pull back the protective Cover (item 1) from the Fibre Output Termination (item 2).
4
Ensure that the rear edge of the Fibre Output Termination is
NOT
chamfered.
5
Ensure the Locating Pin (item 3) on the Fibre Output Termination (item 2) is aligned
with the Locating Hole (item 5) on the Focus Head (item 7).
6
Insert the Fibre Output Termination (item 2) into the Focus Head (item 7).
The Fibre Output
Termination
will enter the Focus Head easily and then stop. A slight
pressure is then required to ensure that the FCMS connectors have fully connected. The
Fibre Output
Termination
is fully inserted when no aluminium on the Fibre Output
Termination
is visible at the entrance to the Focus Head.
7
Tighten the Retaining Clamp screw (item 6) to secure the Fibre Output Termination
(item 2) in the Focus Head (item 7).
8
Position the protective Cover (item1) over the Fibre Output Termination (item 2).
9
If necessary, place a protective cover over the exposed end of the Focus Head
(item 7).