GMC-I Messtechnik GmbH
97
Error while writing the data backup file to
the USB flash drive
There’s not (no longer) enough avail-
able memory space on the storage
medium.
–
In particular in the case of FAT16
formatted USB flash drives: too
many files on the USB flash drive
–
Power consumption of the utilized
USB flash drive exceeds 500 mA.
–
The USB flash drive was discon-
nected during data import.
–
The USB flash drive is defective or
incompatible with the test instru-
ment.
➭
Make sure that a at least 100 MB is
available on the USB flash drive
and delete any data files which are
no longer required.
➭
If the problem persists, save the
data from the USB flash drive to
another storage device and refor-
mat the USB flash drive (FAT32).
➭
Only use USB flash drives with
power consumption of less than
500 mA in combination with the
test instrument.
➭
Make sure that the USB flash drive
isn’t disconnected or moved until
the entire data backup process has
been completed.
➭
If none of these measures results in
improvement, replace the USB
flash drive. A list of tested USB
flash drives is included in section
14.3.
Moving of an object has failed
Moving a test object would lead to an
ID conflict. The ID already exists for this
customer.
➭
Delete the object with duplicate ID.
➭
Select another customer as a relo-
cation target.
Errors During Operation with Barcode or RFID Scanner
–
The scanned barcode is too long.
–
While writing an RFID tag an
attempt was made to write an ID to
the tag with vowel mutations such
as ä, ü or ö, or with special charac-
ters.
➭
Change vowel mutations such as ä
to ae.
➭
Avoid the use of special characters
in the ID.
Error Messages
Possible Causes
Corrective Measures
Summary of Contents for SECULIFE ST PRO