GMC-I Messtechnik GmbH
81
Classification Parameter – IEC 62353
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Selection of Applied Parts for the Test Sequence
➭
Press the “
Classification Parameter
” key.
➭
Go to page 2/2.
➭
Press the “
Applied Parts
” key. The screen shown above ap-
pears.
➭
Touch click operation:
Alternatively, this submenu can be ac-
cessed by touching the “APP Allocation” field.
➭
Select the desired applied part sockets with the “
Ex-
pand Group
” key. A red frame identifies the currently se-
lected sockets, always beginning with socket A. A
socket is added each time the key is pressed. The current
number of selected sockets appears in the field at the bottom
right.
➭
The number of selected sockets can be reduced by
pressing the “
Reduce Group
” key.
➭
After the sockets have been selected, assign the re-
spective type using the “
APP Type
” key. The corre-
sponding icons are allocated to the selected sockets
and the type appears in the field at the bottom left.
➭
After selecting the applied parts and assigning a type,
further groups can be created by pressing the “
Next
Group
” key. An already created group can be edited
after selecting it with the help of the input frame.
Note
The creation of additional groups with the help of the
key is only possible if a type has been assigned to
an already selected group.
Parameter
Setting Options / Meaning
1/2
Standard
VDE 0701-0702, see table below
VDE 0701-0702-EDV, see table above
IEC 60601, 2
nd
edition, see table above
EC 60601, 3
rd
edition, see table above
EC 60601, 3
rd
edition GPA, see previous table
IEC 62353
VDE 0701-0702-VLTG, see following table
VDE 0701-0702-PRCD, see table below
IEC 60974-4, see table below
Protection class
1
PCI, PCII or PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adap.
Adapter: AT3-Adapter
Permanent connection: P1+P2 (only with feature H01)
2/2
Measurement type (MT)
1
Passive
Active
Applied parts
Applied parts: none, B, BF, CF or combinations
Type B (body):
Devices of this type are suitable for both
internal and external patient applications, except for use
in direct proximity to the heart.
The following protection classes are permissible:
I, II, III or devices with internal electrical power supply.
Type BF (body float):
Same as type B, but with type F in-
sulated applied parts
Type CF (cardiac float)
Devices of this type are suitable
for use directly at the heart. The applied part may not be
grounded.
The following protection classes are permissible:
I, II or devices with internal electrical power supply. See
page 81 concerning selection of applied parts for the test
sequence.
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & measurement type
Protection class only (PC)
Measurement type only (MT)
Disabled: No auto-detection:
All classification parameters such as connection, protection
class and measurement type must be entered manually.
2/2
APPs
Type of APP
Expand
Reduce
Next group
group
group