96
GMC-I Messtechnik GmbH
Database Processing Error
–
One of the fields was filled in with
invalid content while processing an
existing database object.
➭
Please be certain to complete all
mandatory fields (identified in red).
➭
If necessary, check your entries to
the fields for invalid special charac-
ters.
–
The ID field was not filled in while
creating a new device.
➭
Fill in the ID field.
–
There’s already an object with the
same ID under the “Customer”
database object.
An incorrect barcode has been
selected.
➭
Assign another ID.
Error while writing the “.secu” file to the
USB flash drive
There’s not (no longer) enough avail-
able memory space on the storage
medium.
–
In particular in the case of FAT16
formatted USB flash drives: too
many files on the USB flash drive
–
Power consumption of the utilized
USB flash drive exceeds 500 mA.
–
The USB flash drive has been dis-
connected during data import.
–
The USB flash drive is defective or
incompatible with the test instru-
ment.
Make sure that at least 100 MB is
available on the USB flash drive and
delete any data files which are no lon-
ger required.
➭
If the problem persists, save the data
from the USB flash drive to another
storage device and reformat the USB
flash drive (FAT32).
➭
Only use USB flash drives with
power consumption of less than
500 mA in combination with the
test instrument.
➭
Make sure that the USB flash drive
isn’t disconnected or moved until
the entire data export process has
been completed.
➭
If none of these measures results in
improvement, replace the USB
flash drive. A list of tested USB
flash drives is included in section
14.3.
Error Messages
Possible Causes
Corrective Measures
Summary of Contents for SECULIFE ST PRO