5.4.1
General Procedure for Creating Test Structures
After selection with the
MEM
key, all setting options for the cre-
ation of a tree structure are made available on three menu pages
(1/3, 2/3 and 3/3). The tree structure consists of structure ele-
ments, referred to below as objects.
Results of measurements/tests can only be saved under structure
element types “Device” or “ME equipment” (
m
edical
e
lectrical
equipment), which are also referred to as “test objects” in the fol-
lowing.
Select the position at which a new object will be added.
➭
Use the
or
key in order to select the desired structure
elements.
➭
If a sublevel exists, you can switch to it by pressing the
key,
or you can open a branch.
➭
The open branch is then closed, or you can switch to the next
higher hierarchical level, by pressing the
key.
Creating a New Object
➭
Scroll to the second menu page (MEM 2/3) with the help of
the
key.
➭
After pressing
NEW
, a new object can be created. Depending
on the current position within the hierarchy, the respectively
available object types are suggested. Depending on the ob-
ject type, you’ll have to enter at least an ID number via the
keyboard. If any of the mandatory entries (identified in red)
have not been completed, an error message appears.
➭
Then press the green checkmark in order to accept the
entered values. The display jumps back up to the higher
hierarchical level.
Move an Object (feature KD01, “Z853S –
SECUTEST DB COMFORT”)
➭
Scroll to the first menu page (MEM 1/3) with the help of the
key.
➭
Select the object to be moved (together with sub-objects) with
the scroll keys.
➭
Scroll to the second menu page (MEM 2/3) with the help of
the
key.
➭
Press the
MOVE
icon.
➭
Using the scroll keys, select the position to which the object is
to be moved and confirm by pressing the green checkmark.
Quick Command: Move Object (feature E01 (touchscreen) and fea-
ture KD01, “Z853S –
SECUTEST DB COMFORT”)
➭
Press and hold the object to be moved in the tree view in the
initial window of a test sequence until the activity bar starts to
blink.
Note
Depending on whether or not finger pressure is applied
for a longer period of time in the customer or location
tree, the device can be “moved” to another customer or
“moved” between locations.
➭
Upon releasing finger pressure, the display is automatically
switched to the database view (MEM), from where you can
proceed to the “Move” menu.
➭
Now select the position with the scroll keys to which the ob-
ject will be moved.
➭
The display is automatically returned to the initial window after
confirming with the green checkmark.
Editing an Object –
Changing the Description or ID Number of a Previously Created
Object
➭
Scroll to the first menu page (MEM 1/3) with the help of the
key.
➭
Select the structure element whose designation will be
changed.
➭
Scroll to the second menu page (MEM 2/3) with the help of
the
key.
➭
Press the
EDIT
icon.
➭
Select the parameter whose description will be changed.
The keyboard appears automatically.
➭
Change the displayed designation and acknowledge your en-
try.
Quick Command: Edit Object (feature E01 (touchscreen) and
feature KD01, “Z853S –
SECUTEST DB COMFORT”)
➭
Press and hold a point in the detail view field in the initial win-
dow of a test sequence until the activity bar starts to blink.
➭
Upon releasing finger pressure, the “Edit” menu for a device/
ME equipment (
m
edical
e
lectrical equipment) opens automati-
cally.
➭
After entering or changing the data, the display is automati-
cally returned to the initial window upon confirming with the
green checkmark.
Summary of Contents for SECULIFE ST PRO