20
GMC-I Messtechnik GmbH
Test Structure, Customer View – Hierarchy of Object Levels
or in Devices with Database Extension, Feature KB01, “Z853R – SECUTEST DB+”
Figure 5 Database Structure as Customer View in Test Instruments with Feature KB01, “Z853R – SECUTEST DB+”
Note
Grayed Out Database Elements
The corresponding elements are grayed out in devices
without enabling for the following options: “extended
database structure”, feature KB01, “Z853R – SECUTEST
DB+” (= property, building, floor, room) or feature KD01,
“Z853S –
SECUTEST DB COMFORT” or medical electri-
cal equipment.
Note
Mandatory Entries
Mandatory entries are identified in red in the entry fields at
the test instrument as well as in the illustrations in Figure
4 and Figure 5.
Note
Hierarchies
The following hierarchies must be complied with:
Room or Floor
must always be subordinate to
Building
.
Devices or ME equipment
(
m
edical
e
lectrical equipment)
must always be allocated to a
Customer
.
Hierarchies and Data Migration
Database objects “Device” and “ME equipment” must always be
subordinate to a Customer. If so-called “legacy data” have been
imported into the test instrument which do not comply with this
rule (e.g. as a result of a firmware update or via the “Restore data-
base” function), Customer objects are generated automatically.
The same applies to database objects “Room” and “Floor”, which
must always be subordinate to a Building. In this case, Building
objects are generated automatically if necessary.
5.2.4
Switching Between 2 Tree Structure Views
(at devices with database extension, feature KB01, “Z853R –
SECUTEST DB+”)
➭
The display can be switched back and forth between the lo-
cation and customer views by repeatedly pressing the
MEM
key.
➭
The database view can be exited by pressing the
ESC
key.
Customer
ID
Designation
Street
Device
ZIP code
City
Database
Manual
ID
Designation
Measurement 1
Measurement 2
Measurement 3
sequence
* Test interval: additionally available for test instruments with database extension, feature KB01, “Z853R – SECUTEST DB+
”
or feature KD01,
“Z853S – SECUTEST DB COMFORT”
** Only with database extension, feature KD01, “Z853S – SECUTEST DB COMFORT”:
additional entry option for ME equipment (
m
edical
e
lectrical equipment) and further parameters such as UDI and applied parts
ME equipment **
UDI = unique device identification
1/6
2/6
3/6
4/6
5/6
6/6
ID
Test interval *
Comment
No. of type B APPs **
Cost center
Mains connection **
Designation
Type
Serial no.
No. of type BF APPs ** Department
Location
Manufacturer
Protection class
No. of type CF APPs ** UDI **
Customer
Customer
Customer MEM