GMC-I Messtechnik GmbH
53
Maximum Permissible Limit Values for Equivalent Leakage
Current in mA
I
GA
Device leakage current
I
EA
Equivalent leakage current
PE Protective conductor
1
For devices with heating power
≥
3.5 kW
2
This limit value isn’t taken into consideration in the DIN EN 62353
(VDE 0751-1) standard.
See section 14.8 for further limit values.
Test Standard
I
GA
I
EDL
VDE 0701-0702
PC I: 3.5 / 1 mA/kW
1
PC II: 0.5
IEC 62353
(VDE 0751-1)
PC II
0.2
2
PC I (PE or parts connected to PE)
1
Permanently connected devices with PE
10
Portable X-ray devices with additional PE
5
Portable X-ray devices without additional PE
2
Devices with mineral insulation
5