14
GMC-I Messtechnik GmbH
4.3
Device Settings
For the purpose of
initial start-up
, we recommend
setting the following basic parameters in the order
shown at the right:
Setup 2/3 > Culture >
Language
(for user interface)
Setup 2/3 > Culture >
Keyboard Layout
(for alphanumeric entries)
Setup 1/3 > System >
Date / Time
(for report generation)
Setup 1/3 > System >
Brightness
(display brightness as %)
Setup 1/3 > Auto Measurements
> 2/2 > Initial Window Style:
Tree or Detail View
Figure 2 Device Settings, Main Menu Level – SETUP Switch Setting
The following parameters are advisable for
maintenance purposes
:
SETUP 3/3 > Test >
Display / Buzzer
(for checking info and warning
displays/signals)
SETUP 3/3 > System Info >
Software Version
for updates (see sec-
tion 13.3) and
Calibration Data
for adjustment, last and next calibra-
tion (see notes on bottom of page 15).
See section 13.3 regarding downloading the latest software version.
Notes Concerning the Inspector Parameter
–
The inspector who has just been “selected” appears in com-
pleted tests as the “Inspector”. None of the test instrument’s
settings are stored specifically for the inspector – all of the set-
tings at the test instrument are stored for the respective
device and are available to
all
inspectors.
–
If an inspector is password-protected
, this only prevents users who don’t
know the password from “selecting” this inspector. There’s
no
pass-
word prompt when the test instrument is started up. The inspector
remains selected even in the event of a power failure – a
(password-protected) inspector can only be unselected by
selecting another inspector.
In order to delete an inspector whose password you don’t
know, it’s sufficient to enter an incorrect password five times
in a row and to confirm the entry each time – a query then
appears as to whether or not the inspector should be deleted.
The inspector to be deleted may not be the same as the cur-
rently selected inspector.
Setup
Display additional menu pages
Menu selection for operating parameters, page 1 of 3
Clear database or display statistics,
Display additional menu pages
Menu selection for operating parameters, page 2 of 3
Add, select, delete or protect inspector
Select and configure a barcode reader
Display additional menu pages
Menu selection for operating parameters, page 3 of 3
Keys with fixed functions
Softkeys = keys with functions that vary
depending on the operating level
Setup 1/3
Setup 2/3
Setup 3/3
Save
only appears if a USB flash drive is plugged in
by scanning the displayed barcode
with password (see note in this regard below)
Set test sequence parameters: amongst others measur-
ing sequences (standard) (see section 10.3)
Parameters for single measurements and test
sequences (see section 4.1.1 and section 6.2)
Menus for using the
Bluetooth
®
interface
(see section 14.4) (only with feature M01)
To submenu for selecting and setting up a printer
(see page 17)
Only with database extension or feature KB01:
manage and import own sequences
To submenu for setting up test reports
(see section 3.5.3)
Query software/hardware version, serial no.,
calibration data, memory occupancy, user’s manual *
* QR code with link to current operating instructions on the web
Culture, date/time, default settings, brightness,
volume ...
(see submenu on following page)
ESC
HELP
MEM
ESC
HELP
MEM
ESC
HELP
MEM