Note
Measurements and measurement series can only be
saved after measurement has been completed. Mea-
sured values can only be added to intermediate buffer
memory during a measurement. Customer, location and
other entries cannot be changed in the memory menu.
These have to be selected directly in the database and
entered or changed.
Note
Please observe the following before storing tests or measure-
ments to the test instrument:
If applicable, the date of recalibration is printed on test
reports, or transmitted to a PC when exporting test data.
For this reason we recommend checking the recalibration
date saved in the test instrument before starting work
with your new test instrument (see page 15).