Setting Measuring Parameters for IPA
Note
The currently selected
single fault
is displayed on
the initial page. In order to change it, it has to be
selected and
set via the measuring parameters
menu.
Test Sequence
➭
Set the rotary switch to the
I
PA
position.
➭
Connect the DUT in accordance with the selected measuring
type.
➭
Connect the applied parts.
➭
Select the measurement type:
– By setting the parameters
or
– Directly via the
Measurement Type
key
➭
Select the respective applied parts by setting the utilized
sockets to “on” and the unused sockets to “off”.
➭
In the case of
direct measurement
P1
, measurement
must be performed with mains plug polarity in both di-
rections. Select the respective polarity to this end by
pressing the
NL/LN
key.
➭
Start the test:
press the
START
/STOP
key.
➭
After each reconnection to the mains, and as soon as
the first test is started, a mains connection test is exe-
cuted.
➭
In the case of measurement type direct APP:
Acknowledge the warning which indicates that line
voltage will be connected to the test socket.
➭
Switch the device under test on.
➭
The measured values are displayed. The measured
value recording symbol shown at the right appears.
Each time this key is pressed, the currently displayed
measured value is saved to buffer memory.
➭
Turn off the device under test.
➭
End the test:
press the
START
/STOP
key.
The save symbol appears (floppy disk showing the
number of measured values stored to buffer memory)
and prompts you to save the measured values to an ID
number.
➭
Read the measured values and compare them with the table
of permissible limit values.
➭
Press the
ESC
key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.
See limit values in section 14.8.
Measuring
Parameter
Meaning
Measurement Type
Suitable for
DUT Connection via
Direct APP
Direct measurement (via test
socket) from the selected APP
socket to all others.
Test socket
Perm. conn. APP
Measurement at the permanently
installed DUT from the selected
APP socket to all others.
Permanent connection
APP – applied parts
A / B / C / D / E / F / G / H / I / K, each individually (via on/off) to all other APPs
Single fault (SFC) – only with direct measurement type
Normal state / N interrupted / PE interrupted
Polarity
Only with direct measurement
type
L/N or N/L
Selection of polarity for mains voltage to the test socket