GMC-I Messtechnik GmbH
95
–
Excessive (touch) voltage at test
probe P1
–
Test probe P1 was contacted
although this is NOT specified for
the selected measurement type.
➭
Determine whether or not use of
test probe P1 is permissible in the
currently selected measurement
type (if applicable on the basis of
the illustration in the help image). If
the test probe was contacted erro-
neously, make sure it’s discon-
nected from the measuring circuit
before restarting the measurement.
Otherwise:
➭
If this problem occurs during a
leakage current measurement
, be
sure to check the DUT for insula-
tion faults between housing and
voltage conducting parts!
➭
If this problem occurs during a
pro-
tective conductor resistance measure-
ment
, check the test setup. No
voltage conducting parts may be
contacted with test probe P1
during this measurement.
➭
If this error occurs frequently de-
spite correct (voltage-free) test
setup, please contact our service
department.
➭
Regardless of the utilized measur-
ing function, the possibility cannot
be ruled out that the
fuse link
for
the probe socket (measurement
with test Pro P1) or one of the fuse
links for the applied part sockets
(only with feature J01) has blown.
Check and replace fuses if neces-
sary.
Error Messages
Possible Causes
Corrective Measures