MAXWELLIAN MODE
==STATISTICAL DGD EMULATION==
WITH CORRESPONDING SOPMD
========================
*PME:EMU ENA#
Start PMD-statistical emulation
(1
st
and 2
nd
order)
*E00# if successful. See Table 2 for
other error code definitions.
*PME:EMU DIS#
Stop PMD-statistical emulation
*E00# if successful. See Table 2 for
other error code definitions.
*PME:EMU AVE?
Query average DGD value setting for PMD
Emulation. Default: 10.0 ps
*dgd# (Units in ps)
Example: *23.5#
*PME:EMU:AVE
dgd
#
Set PMD-emulation average DGD value
Range: 90ps PMDPro = 0.0ps to 30ps
180ps PMDPro = 0.36ps to 60ps
*E00# if successful. See Table 2 for
other error code definitions.
(Units in ps )
*PME:EMU SMP?
Query number of samples for PMD emulation
Range: 1 to 9,999, Default:100
*
smp
#
*100#
*PME:EMU:SMP
smp
#
Set number of samples for distribution
Sample Range: 1 to 9,999
*E00# if successful. See Table 2 for
other error code definitions.
*PME:EMU PMD?
Query current DGD and SOPMD values
during an emulation run.
*23.5,235,8# (Units in ps and ps
²
)
*PME:EMU:CYC
cyc
#
Set number of cycles to run the emulation.
Range: 1 to 99,999
*E00# if successful. See Table 2 for
other error code definitions.
*PME:EMU CYC?
Query PMD-emulation cycle number setting.
Default: 10 cycles
*cyc#
Example: *1000#
*PME:EMU:INT
int
#
Set dwell time per point for emulation.
Interval Range: 1 to 99,999 ms
*E00# if successful. See Table 2 for
other error code definitions.
*PME:EMU INT?
Query dwell time setting.
Default: 100.0 ms.
Interval Range: 1 TO 99,999 ms
Example: *1000# (Units in ms)
SOP CONTROL MODE
===============================
============================
*SOP:RAN ENA#
Begin (enable) random scrambling
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:RAN DIS#
Stop (disable) random scrambling
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:RAN FRQ?
Query random scrambling frequency setting
*123#
Results are in Hz.
*SOP:RAN:FRQ
frq
#
Set frequency for random scrambling, in Hz
Range: 1 to 6000
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:TRI FRQ?
Query triangle scrambling frequency setting
*331.1#
Results are in Hz.
*SOP:TRI:FRQ
frq
#
Set frequency for triangle scrambling, in Hz
Range: 0.01 to 500.0, default 100.0
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:TRI ENA#
Begin (enable) triangle wave scrambling
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:TRI DIS#
Stop (disable) triangle wave scrambling
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:SSS
s1,s2,s3
#
Set S1, S2, and S3
Range for S1, S2 and S3:
−
1.000 to +1.000
*E00# if successful. See Table 2 for
other error code definitions.
*SOP SS1?
Query the set value of S1
Return: *0.463# (Example)
*SOP SS2?
Query the set value of S2
Return: *0.123# (Example)
*SOP SS3?
Query the set value of S3
Return: *
−
0.423# (Example)
*SOP:SSS ENA#
Execute SOP setting. If no Stokes parameters
were entered, or if an incorrect set of Stokes
parameters was entered, executes setting
(1, 0, 0)
*E00# if successful. See Table 2 for
other error code definitions.
*SOP:SSS DIS#
Disables active SOP maintenance, returns to
Stokes setting screen.
*E00# if successful. See Table 2 for
other error code definitions.