shows the DGD and SOPMD, sample number, number of completed cycles, and input
and output DOP. The generated DGD distribution plot shows all points generated so far.
When the emulation is complete, it will match the calculated DGD distribution plot above
it. The DOP vs. DGD graph plots the output DOP vs. DGD for all generated points.
For dwell time values > 1 s, the WSOP function is also available. Click the “WSOP
Enable” button at the bottom left of the screen before starting the emulation to enable
automatic polarization optimization for the worst-case PMD effect at each DGD value.
Please note that the WSOP function is only effective for single polarization signals.
For dwell time values <1 s, the WSOP and scan monitoring functions are disabled. The
monitor plots are replaced by a progress bar:
Document #: GP-UM-PMD-1000-21
Page 70 of 122