*PME:SCN MOD?
Query the scan MODE.
*UNI# For UNI-Directional
*BID# For BI-Directional
See Table 2 for error code definitions.
*PME:SCN STE?
For step trace scan:
Query the DGD step size, in picoseconds.
*34.6#
*PME:SCN:STE
step
#
For step trace scan:
Set the DGD step size, in picoseconds.
Range: 0.1 to 64.3 ps for max. range scan.
Maximum step size is limited by range set.
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN RTE?
For continuous trace scan:
Query the scan rate. Result given in ps/s
*34.6#
*PME:SCN:RTE
rate
#
For continuous trace scan:
Set the SCAN RATE in ps/s
Range: 0.1 to 100.0 ps/s
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:CON
dgd
#
For continuous or step trace scan:
Set the start DGD.
PMDPro will match to DGD value of closest DGD
ID# and calculate corresponding trace endpoint.
Example: *PME:SCN:CON 14#
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN CON?
For continuous or step trace scan:
Query scan start point (DGD) and end point (DGD,
SOPMD)
*min-dgd,max-dgd,max-sopmd#
e.g. *19.99,52.19,1137.9#
*PME:SCN:MBA
mn,mx
#
*PME:SCN:MBB
mn,mx
#
*PME:SCN:MBC
mn,mx
#
*PME:SCN:MBD
mn,mx
#
*PME:SCN:MBE
mn,mx
#
Scan PMD values stored in a memory bank
From memory location mn to mx:
Range for mn/mx: 0 to 99,
mn < mx
MBX = Memory Bank X
X = A, B, C, D or E
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:TYP
bnk,mn,mx
?
Query whether Location Range:
mn
to
mx
of
Memory Bank:
bnk
contains only one type (discrete
(lookup table) or continuous ) of PMD values or a
mixture of types.
*MIX#: Range contains both discrete
and continuous PMD values
*UNI#: Range contains only discrete or
only continuous PMD values
*PME:SCN DIS#
Stop Scan
(Discrete mode DGD/PMD scan or Memory bank
scan)
*E00# if successful. See Table 2 for
other error code definitions.
MAXWELLIAN MODE
==STATISTICAL DGD EMULATION==
========================
*PME:MXW ENA #
Start 1
st
order PMD Maxwellian statistical emulation *E00# if successful. See Table 2 for
other error code definitions.
*PME:MXW DIS#
Stop Maxwellian statistical emulation
*E00# if successful. See Table 2 for
other error code definitions.
*PME:MXW:AVE
dgd
#
Set Maxwellian average DGD value, in ps
Range: 0 to (Max DGD)/2
*E00# if successful. See Table 2 for
other error code definitions.
*PME:MXW AVE?
Query average DGD value setting.
Value given in ps.
*20#
*PME:MXW DGD?
Query current DGD ID# during scan
*23#
*PME:MXW:SMP
sample
#
Set number of samples for Maxwellian emulation
Range: 1 to 999,999
*E00# if successful. See Table 2 for
other error code definitions.
*PME:MXW SMP?
Query sample number setting
*10000#
*PME:MXW:CYC
cyc
#
Set number of cycles to run the Maxwellian
emulation
Range: 1 to 99,999
*E00# if successful. See Table 2 for
other error code definitions.
*PME:MXW CYC?
Query cycle number setting
*1000#*
Document #: GP-UM-PMD-1000-21