Document #: GP-UM-PMD-1000-21
Page 49 of 122
Drivers installed:
•
LabVIEW Run-Time Engine 2012 SP1 - (32-bit)
Operating system: Windows 8; Windows 7; Windows Vista; Windows XP (SP3)
32-bit; Windows Server 2008 R2 64-bit; Windows Server 2003 R2 32-bit
•
NI-VISA Run-Time Engine 5.3
Operating system: Real-Time OS; Windows 7 64-bit; Windows 7 32-bit;
Windows Vista 64-bit; Windows Vista 32-bit; Windows XP; Windows XP
Embedded; Windows Server 2008 R2 64-bit; Windows Server 2003 R2 32-bit.
•
FTDI USB Driver (requires additional steps- see USB Setup section). See
Readme file for USB driver for version information.
4.3.3
RS-232 Connection
The RS-232 connector on the rear panel of the PMDPro is a DB9 male connector. Use a
straight connection RS-232 cable (DB-9 female to female) to connect the PMDPro to the
RS-232 port of a computer. To ensure proper communication, use a serial cable with
direct pin-to-pin connected wires (see Figure 23) at both ends of the cable.
6
7
8
9
1
2
3
4
5
Figure 23 RS-232 connector on rear panel of PMDPro
RS-232 command notes:
1.
RS-232 port uses asynchronous framing, 8 data bits, no parity bit, and 1 stop bit.
2.
RS-232 baud rate: 28800 bps.
4.3.4
USB Connection
USB communication requires installation of an additional driver. After the control
program is installed and the computer is restarted, connect the PMDPro to the computer
with a USB cable and power on the PMDPro. This should initiate installation of the USB
driver.
GND
RXD
TXD