arrows to increment the selected digit. The range for average DGD is 0.0 to
30.0 ps for the 90-ps PMDPro and 0.0 to 60.0 for the 180 ps PMDPro. The
ENTER key selects the displayed value.
c.
SAMPLE NUM:
This option specifies the number of samples in the distribution. Use the left
and right arrows to move between digits and the up and down arrows to
increment the selected digit. The range is 1 to 9999. The ENTER key selects
the displayed value.
d.
CYCLE NUM:
This option sets the number of times to run the emulation. The cycle range is
1-99999. Use the left and right arrows to move between digits and the up and
down arrows to increment the selected digit. The ENTER key sets the number
of cycles to the displayed value.
e.
INTERVAL:
This option sets the dwell time per sample, in ms. Use the left and right
arrows to move between digits and the up and down arrows to increment the
selected digit. The range is 1 to 9999 ms. The ENTER key selects the
displayed value.
6.
SCAN PMD MEM BANK:
Scan through a user generated sequence of PMD values. There are 5 memory
banks (labeled A-E), each with 100 storage locations; therefore, up to 5 sequences
of PMD values can be stored and used for this scan. See the STORE menu section
for instructions on storing sequences of PMD values. The PMDPro will scan
through the memory locations in the selected bank in order, starting at the selected
beginning point and ending at the selected end point.
The setup screens for a memory bank scan are as follows:
a. START (SCN BNK)
↓
b. SCAN PMD-MEM BNK: A
c. BGN A:00
ID#0039
↑
d. END A:99
(Contin)
↓
e. INTERVAL: _ 1000 ms
↑
f. CYCLES: _ _10
Use the up and down arrows to move between scan setup options and the
ENTER key to access the setup screen for the selected option.
Document #: GP-UM-PMD-1000-21
Page 27 of 122