d.
CYCLE NUM:
This option sets the number of times to run the emulation. The cycle range is
1-99999. Use the left and right arrows to move between digits and the up and
down arrows to increment the selected digit. The ENTER key sets the number
of cycles to the displayed value.
e.
INTERVAL:
This option sets the dwell time per sample, in ms. Use the left and right
arrows to move between digits and the up and down arrows to increment the
selected digit. The range is 1 to 60000 ms. The ENTER key selects the
displayed value.
B.
PMD EMULATION
The setup screens for first and second order PMD emulation are as follows:
Use the up and down arrows to move between scan setup options and the
ENTER key to access the setup screen for the selected option.
a.
START:
This option starts the emulation. Use the other options to perform the setup,
then return to option a) to begin generating PMD values.
During the emulation, the screen displays:
Top line: current DGD and SOPMD
CY#: number of completed cycles.
SAM#: current sample number.
e. INTERVAL: _2000 ms
↑
(Rng: 1 ms to 9999 ms)
c. SAMPLE NUM: _1000
↑
d. CYCLE NUM: _ _ _10
↓
9.72 ps
436.75 ps2
CY#: 0
SAM# 80
a. START (PMD Emula)
↓
b. AVERAGE PMD: 10.0ps
b.
AVERAGE PMD:
This option specifies the average DGD value for the Maxwellian distribution.
Use the left and right arrows to move between digits and the up and down
Document #: GP-UM-PMD-1000-21