Measurement and Analysis
140
Vivid i User’s Manual
2378958-100 Rev. 02
The formula for this example is as follow:
My LIMP = (MCO-AV ET)/AV ET
1.
In the user-defined folder (e.g.”My folder”), select the last
measurement created (e.g. AV ET).
2.
Double-click
(Name)
in the last line in the
Parameter list
in
the
Measurement menu sheet
.
3.
Enter the name for the formula (e.g. My LIMP).
4.
Select .
The
Edit formula window
is displayed.
Figure 6-12: The Edit formula window
5.
Select “
(
“ from the
Operators drop-down menu
.
6.
In the
Doppler drop-down list
, select
MCO [My Folder, MCO]
.
Make sure to select the measurement located in the user
defined folder (e.g. “My Folder”).
7.
Select “
-
“ from the
Operators drop-down menu
.
8.
In the
Doppler drop-down list
, select
AV ET [My Folder, AVET]
.
9.
Select “
)
“ from the
Operators drop-down menu
.
10. Select “
/
“ from the
Operators drop-down menu
.
11. In the
Doppler drop-down list
, select
AV ET [My Folder, AVET]
.
The Formula line should display:
({MCO}-{AVET})/{AVET}
.