Archiving
194
Vivid i User’s Manual
2378958-100 Rev. 02
4.
In the
Remote archive setup area
enter:
• The IP address for the remote archive
• The name of the remote archive
5.
Press
Save settings
and reboot the system.
Direct connect with online acquisition
Data management
Scanner’s dataflow configuration
1.
Select the
RemoteArch-RemoteHD
dataflow as default
dataflow.
When saving the study on the scanner, the images are
transferred from the scanner’s image buffer to the
harddrive of the EchoPAC PC. Patient demographics,
measurements and reports are transferred on the fly when
entering the information on the Vivid
i
.
Review on the EchoPAC PC workstation
1.
Select the
LocalArchive-Int.HD
dataflow on the
EchoPAC PC (can be configured as default dataflow).
2.
Press
Archive
and select the patient to be reviewed.
CAUTION
Do NOT open a study on the EchoPAC PC workstation before the
study is closed on the scanner.