Measurement and Analysis
120
Vivid i User’s Manual
2378958-100 Rev. 02
See the
Status bar
to get the next step
to perform.
6.
Press
SET
to anchor the start point of the measurement.
7.
Trackball the cursor to draw the length.
8.
Press
SET
to anchor the second point.
9.
Drag the cursor with the
Trackball
to outline the area of
interest.
The measurement
display color on the
2D image changes
from green to red af-
ter completion of
the measurement.
The current area, circumference and Area/Length Volume
(Va-l) values are displayed in the
Measurement result table
(see Figure 6-1) and are instantaneously updated when
moving the cursor.
10. Press
SET
to complete the measurement.
The measurement results including Vmod (Simpson) are
displayed in the
Measurement result table
(see
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
11. To assign a label to the measurement, see page 115.
12. Repeat steps 5 through 10 to make additional volume
measurements.
2D Depth measurements
The measurements described in this section enable depth
measurement from the probe to a selected point.
To perform a depth measurement:
1.
Generate the 2D image.
2.
Press
FREEZE
to stop the cineloop.
3.
Press
MEASURE
on the Control Panel.
See the
Status bar
to get the next step
to perform.
4.
Press the assignable
POINT
to select the depth
measurement function.
5.
Trackball the cursor to the position to measure.
The current distance from the probe is displayed in the
Measurement result table
and is instantaneously updated
when moving the cursor.
The measurements
displayed on the 2D
image and the cor-
responding results
are numbered.
6.
Press
SET
to anchor the point.
The depth value (cm) is displayed in the
Measurement
result table
.
To Purchase, Visit
Summary of Contents for Vivid i
Page 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Page 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Page 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Page 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Page 387: ......
Page 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...