Archiving
198
Vivid i User’s Manual
2378958-100 Rev. 02
A scanner and a DICOM server in a
network
In this scenario the Vivid
i
is configured to work with a DICOM
server in a network environment. Images are first saved on the
local image buffer on the scanner. At the end of the
examination the images are sent to the DICOM server via a
DICOM spooler.
The DICOM server dataflow supported are:
•
Local Archive - Int HD/DICOM Server
: the local archive is
used for patient archiving. Images are stored to the internal
harddrive and to a DICOM server.
•
Remote Archive - Remote HD/DICOM Server
: a remote
database is used for patient archiving. Images are stored to
a network image volume and to a DICOM server.
•
Worklist/Local Archive - DICOM Server/Int HD
: search in
a DICOM Modality Worklist, the patient found is copied into
local database. The patient information and the
examination results are stored to the local database.
Images are stored to a DICOM server and to an image
volume on the local harddrive.
•
Worklist/Remote Archive - DICOM Server/Remote HD
:
search in a DICOM Modality Worklist, the patient found is
copied into a remote database. The patient information and
the examination results are stored to a remote database.
Images are stored to a DICOM server and to an image
network volume as pure DICOM in both locations.
•
Query/Retrieve
: retrieve images from a DICOM server
based on query parameters.
This scenario requires that the scanner is configured to be
connected to the DICOM server as described below.
Scanner’s connectivity configuration
The scanner’s TCP/IP settings must be configured as
described in "Scanner’s TCP/IP settings" on page 193.
In addition, to work against the DICOM server the following
information has to be entered in the scanner:
•
The DICOM server IP address
•
The DICOM server port number
To Purchase, Visit
Summary of Contents for Vivid i
Page 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Page 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Page 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Page 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Page 387: ......
Page 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...