Scanning Modes
Vivid i User’s Manual
87
2378958-100 Rev. 02
7.
Repeat steps 4. and 5. to change the angle of the solid
cursor line if necessary.
The M-Mode area
of the display up-
dates as the
M-Mode signal is
constructed.
8.
Press
TRACKBALL
to activate scrolling control on the
trackball.
9.
Use the
trackball
to scroll through the data acquired at that
location. The M-Mode display will vary accordingly.
Optimizing M-Mode
Refer to page 291
about creating pre-
sets.
The use of preset gives optimum performance with minimum
adjustment. If necessary, the following controls can be adjusted
to further optimize the M-Mode display:
• Adjust
Horizontal sweep
to optimize the display resolution.
• Adjust
Gain
and
TGC
controls to adjust the range to be
imaged.
Except for Contour,
all the controls list-
ed in the optimizing
M-Mode section
will also affect the
2D image.
• Use
the
Frequency
(move to higher frequencies) or the
Frame rate
control (move to lower frame rate) to increase
resolution in image.
•
Use the Frequency (move to lower frequency) to
increase penetration
.
• Adjust
Focus
to move the focal point(s) around the region
of interest in the M-Mode display.
• Adjust
Dynamic range
to optimize the useful range of
incoming echoes to the available greyscale.
• Adjust
Compress
and
Contour
to further optimize the
display.
• Adjust
Reject
to reduce noise while taking care not to
eliminate significant low-level diagnostic information.
To Purchase, Visit
Summary of Contents for Vivid i
Page 110: ...Scanning Modes 100 Vivid i User s Manual 2378958 100 Rev 02...
Page 162: ...Purposely Left Empty 152 Vivid i User s Manual 2378958 100 Rev 02...
Page 288: ...Probes 278 Vivid i User s Manual 2378958 100 Rev 02...
Page 350: ...User maintenance 340 Vivid i User s Manual 2378958 100 Rev 02...
Page 387: ......
Page 388: ...To Purchase Visit Avobus com or call 1 800 674 3655...