Contrast Imaging
Vivid i User’s Manual
107
2378958-100 Rev. 02
The Diff control de-
creases the frame
rate and the number
of focal zones when
turned on.
Diff on/off
Affects the level of reverberation in the image. The
reverberation in the image is reduced when Diff control is
turned on.
DDP (Data Dependant Processing)
Performs temporal processing, which reduces random noise
without affecting the motion of significant tissue structures.