LV Contrast overview
Figure 5-1: The LV Contrast acquisition screen
LV Contrast controls
LV Contrast assignable controls
Width
Controls the size and angular width of the image sector. A
smaller angle generally produces a scan with a higher frame
rate.
Frequency
Enables the adjustment of the probe's operating frequency. A
higher frequency gives better resolution.
Focus
Changes the location of the focal point. A triangular marker on
the depth scale along the image sector indicates the position of
the focal point.