Archiving
Vivid i User’s Manual
205
2378958-100 Rev. 02
2.
Select one of the following:
•
Save
: the supported measurements are stored as
DICOM SR.
•
Ignore
: no measurements are saved as DICOM SR.
•
Cancel
: cancels the end exam process.
Adjusting the Search criteria
When selecting a DICOM Worklist dataflow or Query/Retrieve,
search criteria can be set for the system to use when searching
the database.
1.
Press
CONFIG
and log on as administrator.
2.
Select the
Connectivity
category and
Dataflow
subgroup.
The
Dataflow sheet
is displayed (see Figure 8-38).
3.
Select a DICOM Worklist dataflow or the Query/Retrieve
dataflow.
4.
Select the Worklist or Query/Retrieve device in the
Selected devices pane
and press
Properties
.
The
Properties window
for the selected DICOM device is
displayed.
5.
Press
Search criteria
.
The
Search criteria window
is displayed.
6.
Select a Search criteria from the
Select tag pull-down
menu
.
7.
Enter a value if required or leave blank if not to be used.
This entry is case sensitive and must match exactly.
8.
Press
Add to list
.
9.
Press
OK
to close the Search criteria window.