Chapter 10 - Care and Maintenance
10-17
P R E L I M I N A R Y
10-6-3
Outlet Test - Wiring Arrangement - USA and Canada
Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet
tester and noting the combination of lights that are illuminated. Any problems found should be reported
to the hospital immediately and the receptacle should not be used.
NOTE:
No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the
Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this
should be suspected as a possible cause and the outlet wiring should be visually inspected.
Figure 10-263 Typical Alternate Outlet Test
1.
Correct Wiring
4.
Open Neutral Wire
2.
Open Ground Wire
5.
Hot and Ground Reversed
3.
Reversed Polarity
6.
Open Hot Wire