Chapter 8 - Replacement Procedures
8-205
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-9-4
Software Installation Procedure - General Overview
The Venue™ ultrasound scanner software loading options provide the ability to install the Windows
operating system and the Venue™ application software.
The complete installation procedure can be performed using the Touch Screen only. An on-screen
(virtual) keyboard is available whenever text input is required. I
When installing the Windows operating system software, after replacement of the SSD module, it is
necessary to format the hard disk.
Be aware that this will perform a full format of the Hard Disk and
all patient data will be lost
.
Normally, the Venue™ system is supplied with the software already installed. In some cases, it is
necessary to re-install the software.
NOTE:
After performing a software installation, all logs are deleted. It is recommended that these
should be recorded should they be needed in the future.
If you are upgrading the system software from a previous version, it is necessary to contact your local
OTR department to receive an appropriate software password. Make sure you specify the system serial
number located on the Venue™ unit.
8-9-4-1
Preparation for Software Installation Procedures
NOTE:
Make sure that the required Software Option keys are available prior to commencing the installation
process. Do not use the Software Option keys from a previous system version.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the "Copy" option
(for instructions, refer to the Venue™ User Manual on your Document Media).
NOTE:
It is recommended that Disk Management be executed using the "move" option to a network
repository.
WARNING
WARNING
DO NOT ATTEMPT TO INSTALL SOFTWARE THAT WAS NOT DESIGNATED FOR YOUR
VENUE™ UNIT. ATTEMPTING TO INSTALL UN-APPROVED SOFTWARE WILL CAUSE
IRREVERSIBLE DAMAGE TO HARDWARE AND SOFTWARE!
NOTICE
IMPORTANT Since neither the System Serial Number or the Computer Name can be changed after the
software installation procedure is completed, it is important to make sure that when prompted, the
correct information is inserted during the installation procedure.
WARNING
WARNING
WHILE THE SOFTWARE INSTALLATION PROCEDURE IS DESIGNED TO PRESERVE
DATA, YOU SHOULD SAVE ANY PATIENT DATA, IMAGES, SYSTEM SETUPS TO
BACKUP MEDIA BEFORE DOING A SOFTWARE INSTALLATION.
NOTICE
IMPORTANT Before performing any Venue™ software installation procedure, it is mandatory to
backup the Archive.
CAUTION
THIS PROCESS FORMATS THE HARD DRIVE (WHEN SELECTION OF THIS OPTION IS
APPLICABLE) - MAKE SURE TO SAVE ALL THE REQUIRED PRESETS, SETTINGS AND
PATIENT DATA!
WARNING
WARNING
REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
BEFORE STARTING THE UPGRADE PROCEDURE.