Connecting the airway module
1.
Check that the airway module is installed.
2.
Check that the airway adapter connections are tight and that the
adapter is correctly installed.
3.
Check that the water trap container is empty and properly
attached.
Empty the water trap container whenever the container is more
than half full. Under normal conditions, the D-fend Pro container
fills in 24 hours.
Before connecting the exhaust line to the sample gas
outlet on the airway module, ensure the other end is
connected to the sample gas return port on the
anesthesia machine. Incorrect connections may cause
patient injury.
4.
Connect a line from the airway module sample gas outlet to the
sample gas return port on the anesthesia machine.
5.
Attach the gas sampling line to the sampling line connector on
the water trap.
6.
Turn the system on. The system does a series of automated self
tests. The automatic agent identification is activated.
7.
Connect the sampling line to the airway adapter. Take the gas
sample as close to the patient’s airway as possible. Position the
adapter’s sampling port upwards to prevent condensed water
from entering the sampling line.
1
2
3
AC.24.064
1.
Airway module
2.
Gas sampling line
3.
Airway adapter with sampling line connector
Figure 6-3 • Airway gases setup with airway module
WARNING
Carestation
™
620/650/650c (A1)
6-6
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...