Auto limits
Use the
Auto Limits
menu to quickly set alarm ranges for ‘MV’, ‘TV’,
and ‘EtCO2’ during mechanical ventilation.
Setting auto limits
1.
Select
Auto Limits
.
The menu shows the current measured values and the
proposed low and high alarm limits.
2.
Check the proposed parameters.
•
Select
Confirm
to use the proposed low and high alarm
limits.
•
Select
Cancel
to leave the alarm limits unchanged.
•
Select
Case Default Limits
to set the alarm limits to the
case default limits.
The proposed low and high alarm limits are shown in highlighted
text. The alarm limits that are not highlighted do not change.
Note
3 Operation
2076152-001
3-15