Message
Priority
Cause
Action
Backup Mode
active
Low
No spontaneous breaths in set period
of time (Backup Time) since starting
PSVPro
™
mode.
Select a new ventilation mode. The
number of consecutive patient triggered
breaths reaches the Exit Backup
setting.
Breathing system
loose
Low
The breathing system is not latched.
Lift the lower breathing system and
ensure it latches. Close breathing
system door.
Calibrate
(remove) flow
sensors
Low
Flow sensor calibration failure.
Calibrate the flow sensors.
Calibrate O2
sensor
Low
Calibration failure or measured O2 is
greater than 110%.
Calibrate the O2 cell. Replace O2 cell if
calibration fails.
Calibrate, dry, or
replace flow
sensors
Low
Patient volume mismatch occurred
during the last case.
Calibrate, dry or replace the flow
sensors. Start a new case.
Confirm to
shutdown
High
On/Standby button is pushed during a
patient case.
Select Confirm within 10 seconds and
the system will shut down. If Confirm is
not selected within 10 seconds, the
system reverts to the screen showing
before the On/Standby button was
pushed.
Cannot monitor
Air pipeline
Medium
Air pipeline pressure is invalid.
Check pipeline supply pressure.
Cannot monitor
O2 pipeline
Medium
O2 pipeline pressure is invalid.
Check pipeline supply pressure.
Cannot monitor
N2O pipeline
Medium
N2O pipeline pressure is invalid.
Check pipeline supply pressure.
Check circuit
connections
Medium
Breaths detected in the circle circuit
while non-circuit is selected.
Check circle circuit connections and
settings.
Check D-Fend
Medium
Water trap not attached.
Check that the water trap is properly
attached to the airway module.
Check flow
sensors
Medium
System has detected an improper flow
pattern in the breathing circuit.
Ensure the flow sensors are connected
correctly.
Check sample
gas out
Medium
Possible blockage in airway module
sample gas outlet.
Check for blockage in the airway
module sample gas outlet. Remove
blockage.
Circuit leak
Medium
Vent TVexp is less than half of vent TV
insp for at least 30 seconds.
Check for leaks in the patient circuit.
Calibrate flow sensors. If problem
persists, replace flow sensors.
Circuit leak
silenced
Low
Setting on Alarm Setup menu. Vent
TVexp is less than 50% of vent TVinsp
for at least 30 seconds and Leak Audio
is changed from On to Off.
Check the patient. Make sure Leak
Audio Off is the appropriate setting for
the case. Check for leaks in the patient
circuit. Calibrate flow sensors. If
problem persists, replace flow sensors.
Circuitry > 75C.
Shutdown
possible.
Medium
Power supply temperature exceeds 75
degrees C.
Shut down system as soon as possible.
Then, check cooling fans and filters.
Cooling fan
needs service.
System OK.
Medium
Fan reporting error.
Shut down system as soon as possible.
Then, check cooling fans and filters.
Carestation
™
620/650/650c (A1)
7-6
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...