Message
Priority
Cause
Action
Service
calibration
advised
Low
Calibration data is corrupt.
Contact an authorized service
representative.
System leak?
Low
Leak detected between ventilator and
patient circuit.
Check for leaks in the breathing system.
System shutdown
in < 5 minutes
High
Remaining battery power is between
zero and five minutes.
Plug in the power cable.
Touchscreen
failure
Medium
Touchscreen interface is not working.
Contact an authorized service
representative.
Turn off gas flow Medium
The gas flow is greater than 300 ml in
Standby.
Turn off the gas flow.
Turn power Off
and On for self
tests
Low
System has been operating for longer
than 12 hours without a power-up self
test.
Turn power off and back on between
cases to perform a self test.
TV not achieved Low
Measured tidal volume is less than set
tidal volume.
Check for leaks in the patient circuit.
Check for leaks in the breathing system.
TVexp high
Medium
TVexp is greater than TVexp high
alarm limit (for nine breaths).
Set the alarm limits appropriately.
Check ventilation settings.
TVexp low
Medium
TVexp is less than TVexp low alarm
limit (for nine breaths).
Set the alarm limits appropriately.
Check ventilation settings.
Unable to drive
bellows
Low
Bellows is collapsed.
Check the drive gas. Increase fresh gas
flow (or push the O2 flush button) to fill
the bellows.
Ventilate
manually!
High
Software or hardware failure prevents
mechanical ventilation. No pressure,
no flow, and no volume monitoring
from ventilator.
Use a manual bag to ventilate the
patient or use an alternate machine.
Shut down the system as soon as
possible and contact an authorized
service representative.
Ventilate
manually!
Medium
Software or hardware failure prevents
mechanical ventilation.
Use a manual bag to ventilate the
patient or use an alternate machine.
Shut down the system as soon as
possible and contact an authorized
service representative.
Ventilator has no
drive gas
High
Drive gas supply is not sufficient to
mechanically ventilate.
Check drive gas supply. Use a manual
bag to ventilate the patient until drive
gas supply is restored.
Vol and Apnea
monitoring off
Low
Non circle circuit is selected.
Auxiliary gas is turned on.
Message will clear when circle circuit is
selected.
Vol vent only. No
PEEP or PSV.
Medium
Manifold pressure error. Pressure
control unavailable. Bag/Vent switch is
in Vent and running PCV, PSVPro
™
,
SIMV PCV, CPAP + PSV, PCV-VG, or
SIMV PCV-VG mode.
Use volume control ventilation mode.
Shut down the system as soon as
possible and contact an authorized
service representative.
Vol vent only. No
PEEP or PSV.
Low
Manifold pressure error. Pressure
control unavailable. Bag/Vent switch is
in Vent and running VCV or SIMV VCV
mode, Bag/Vent switch is in Bag, or
Non-circle circuit or ACGO is selected.
Use volume control ventilation mode.
Shut down the system as soon as
possible and contact an authorized
service representative.
Volume sensors
disagree
Low
TVexp is greater than TVinsp for six
breaths.
Calibrate the flow sensors. Replace the
flow sensors if the message does not
clear.
7 Alarms and troubleshooting
2076152-001
7-11
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...