Expiratory breathing resistance in bag mode*
l/min
kPa
cmH2O
Absorber canister installed
5
0.057
0.57
30
0.247
2.47
60
0.560
5.60
Absorber canister removed
5
0.057
0.57
30
0.247
2.47
60
0.560
5.60
*Values include patient circuit tubing and Y-piece 0.065 kPa (0.65 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations affect
resistance.
Pressure flow data (APL valve completely open)
Flow (l/min)
APL pressure cmH2O
(dry)
APL pressure cmH2O
(wet)
Open
0.978
0.996
3
1.07
1.08
30
1.46
1.49
60
2.20
2.31
70
2.65
2.81
Gas scavening
Passive scavenging
Positive pressure relief
10 cmH2O
Negative pressure relief
0.3 cmH2O
Outlet connector
30 mm male taper ISO
Active scavenging
Disposal system
type
Outlet
connector*
Flow range
Disposal
connection
High flow, low
vacuum
BSI 30 mm
threaded
50 to 80 l/min
ISO 1 H
11 Specifications and theory of operation
2076152-001
11-13