AC.24.288
2.
Set the ACGO switch to the ACGO position.
Fresh gas flows through the ACGO port.
Fresh gas oxygen concentration is displayed on the screen if the
system has the airway module option.
3.
Set the alarm limits to clinically appropriate settings.
4.
To stop fresh gas flow through the ACGO port, set the ACGO
switch to the circle circuit position.
Displayed fresh gas oxygen concentration may not
reflect FiO2 during spontaneous breathing or in
rebreathing circuits. Use an external O2 monitor if using
a rebreathing circuit on ACGO.
WARNING
Carestation
™
620/650/650c (A1)
3-38
2076152-001