Electrical problems
If a circuit breaker opens frequently, do not use the
system. Contact an authorized service representative to
repair the system.
.
Symptom
Problem
Solution
Mains indicator is not on.
The electrical power cable is not
connected.
Connect the power cable.
The power cable is damaged.
Replace the power cable.
The electrical socket the power cable
connects to has no power.
Use a different electrical socket.
An internal fuse is open.
Contact an authorized service
representative to repair the system.
The power management board does not
work. The cables between mains and the
power management board or the power
management board and the anesthesia
control board are disconnected.
Contact an authorized service
representative to repair the system.
One electrical outlet does
not have power.
The outlet circuit breaker is off.
The electrical outlet is damaged.
Turn the circuit breaker on.
Contact an authorized service
representative to repair the system.
A circuit breaker opens
frequently.
Equipment connected to the outlet(s) uses
more current than the circuit breaker
rating.
Use a different power supply for some of
the equipment.
The equipment connected to the outlet
has a short.
Contact an authorized service
representative to repair the system.
Tec
™
6 Plus vaporizer has
no power.
Not plugged into outlet.
Connect power cable.
The outlet circuit breaker is off.
Turn the circuit breaker on.
The real time clock on the
anesthesia display does not
maintain accurate time and
date.
The battery in the anesthesia display
needs to be replaced.
Contact an authorized service
representative to repair the system.
Audible alarm sounds. The
system function stops. The
screen is blank.
Internal power converter failure.
Turn the system off. Contact an authorized
service representative to repair the system.
Display does not function.
Display is broken.
Contact an authorized service
representative to repair the system.
Task light does not function Task light is broken.
Contact an authorized service
representative to repair the system.
On/Standby switch does not
work
System does not have electrical or battery
power.
On/Standby switch is broken.
Confirm power cable is connected.
Contact an authorized service
representative to repair the system.
Touch screen does not
function.
Touch screen is locked, dirty or broken.
Make sure there is no liquid or other dirt on
the touch screen and that the screen is
unlocked. Contact an authorized service
representative to repair the system if the
problem continues.
WARNING
Carestation
™
620/650/650c (A1)
7-18
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...