Turning on the system
1.
Plug the power cord into an electrical outlet. Make sure the
mains indicator light is on.
•
The mains indicator is lit when AC power is connected.
•
Battery is charging if it is not already fully charged.
1
2
AC.24.020
1.
Mains indicator
2.
On/Standby switch
Figure 3-1 • Mains indicator and
On/Standby switch
2.
Check that the breathing system is properly connected.
Do not turn on the system with the right-hand
(inspiratory) port plugged.
3.
Push the On/Standby switch for 1 second to turn on the system.
The Total flow indicator shows total fresh gas flow and the
display shows the power-up screen.
The system does a series of automated self tests.
In case of emergency, during power-up, O2 and agent delivery is
possible in bag mode.
4.
Perform a
Full Test
before the first case of the day.
5.
Perform a preoperative checkout before each case. See the
"
" section.
The system must perform a power-up self test after 12 hours of
remaining on. If the system has been on longer than 12 hours
without a power-up self test, the ‘Turn power Off and On for self
tests’ alarm occurs. Turn the power off and then back on between
cases to resolve the alarm.
CAUTION
Note
Carestation
™
620/650/650c (A1)
3-4
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...