ACGO display
A
C
.24.1
19
-20
100
0
20
40
60
ACGO in use: Ventilator Off
Turn off ACGO to use circle circuit.
[ VCV ]
TV
[ 500 ]
RR
[ 12 ]
PEEP
[ Off ]
43
14
11
0.5
1
5
10
15
1
3
2
4
P-ACGO cmH2O
40
0
70
140
P-ACGO
5
6
FGO2 %
Total Flow
4.6
32
1.3 2.4 0.9
7
P-ACGO
8
1.
Split screen
Shows P-ACGO, gas flow values, compliance, trends and optional ecoFLOW
information. For example: P-ACGO.
2.
Waveform fields
When pressure waveform is selected, displays the measured values for pressure
through the ACGO port.
3.
Measured values fields
When pressure waveform is selected, displays the meassured values for pressure
through the ACGO port.
4.
ACGO status
Shows that the ACGO switch is set to the ACGO position.
5.
Status field
Shows the ACGO status and alarm or status messages. For example: ACGO in use:
Ventilator Off.
6.
Message fields
Displays ventilation mode or messages. For example: Turn off ACGO to use circle
circuit.
7.
Ventilator quick keys
Displays preset ventilation parameters. For example: VCV, TV, RR, and PEEP.
8.
Total flow
Shows information for Total Flow and FGO2 %.
Figure 2-15 • ACGO typical full view
Carestation
™
620/650/650c (A1)
2-18
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...