Battery information
The system is not a portable unit. A sealed lead acid battery supplies
backup power in the event of a power failure.
•
Capacity to support full system functionality and ventilation for
90 minutes when fully charged.
•
The system functions to specifications through the transition to
battery power. The system will automatically switch over to the
battery supply when adequate system supply cannot be
maintained from the mains input.
•
The system functions to specifications while the battery is
recharging.
•
Electrical outlets (if provided) will not function during a power
failure.
•
Electrical outlets (if provided) will not function on battery power.
Only trained service representatives are to replace the battery.
Batteries must be disposed of in accordance with applicable
regulatory requirements in effect at the time and place of disposal.
Contact a trained service representative to disconnect
the battery if the equipment is not likely to be used for an
extended time.
WARNING
Carestation
™
620/650/650c (A1)
11-10
2076152-001